Wireless reference implant and communication methodology to assess and investigate RF safety and pTx mitigation strategies for AIMDs

Berk Silemek, Frank C. Seifert, Bernd Ittermann, Lukas Winter · Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition · 2023

Parallel transmission (pTx) systems can substantially improve the RF-safety of active implantable medical devices (AIMDs) in MRI. Here, a wireless reference implant is presented to test pTx mitigation strategies of RF induced implant heating. It is demonstrated that the proposed hardware and communication workflow can measure the sensor Q matrix and use this information to mitigate RF induced heating. The proposed setup enables conceptualization and further testing of a safety strategy relying on an implant communicating with a pTx capable MRI to improve RF safety without major restrictions in MR imaging performance.