Report of METL Mutual Inductance Level Sensor Development for Use in Liquid Metals – FY2023
Edward Kent, M. Weathered, Jochen Rein, D. Kultgen, C. Grandy · 2023
A robust mutual inductance level sensor (MILS) for use in high temperature liquid metals has been developed at Argonne National Laboratory (ANL). The current design utilizes mineral insulated cables with a 300-series stainless steel sheath, magnesium oxide insulation, and a single copper conductor. Two coils are wrapped on a common core made of a 300-series stainless steel tube. One coil is energized with an alternating current (AC) source, and this electromagnetically couples with the second coil to generate an induced voltage. The voltage is measured to determine the mutual induction between the coils, and the mutual inductance measurement can be used to determine the level of a nearby liquid metal volume. The MILS can be located in an isolating thimble that is fully sealed to a vessel containing liquid metal for ease of maintenance and replacement. When the primary coil is energized with the AC source, the coil not only electromagnetic (EM) couples with the second coil, but also the surrounding liquid metal. The EM coupling with the surrounding liquid metal reduces the EM coupling with the second coil, producing an inverse-linear relationship between liquid metal level and secondary coil voltage. This has all been demonstrated at the Mechanisms Engineering Test Loop (METL) liquid sodium facility at ANL. The most current iteration of the MILS system is the MILS-MK-II. This sensor system has been commissioned in a non-sodium test stand where calibrations were performed using a sodium analog. The calibrations proved to be highly linear and repeatable. The MILS-MK-II has been installed in the METL expansion tank where high temperature sodium tests have been performed. The MILS-MK-II has been calibrated against a known standard at temperature of 300°C, and the calibrations have proved to be highly linear and repeatable. The calibrated MILS-MK-II has been in operation in the METL facility for several thousands of hours at temperatures around 300°C. The experimental data has been used to develop and validate electromagnetic finite element models in COMSOL Multiphysics, and now these models can be used to advance the development of the sensor system. Next steps will include temperature compensation to allow for operation at various temperatures. Additionally, efforts to incorporate a self-calibration methodology are underway.