Operational testing of system for automatic sleep analysis

Peter Kellaway · NASA STI Repository (National Aeronautics and Space Administration) · 1972

Tables on the performance, under operational conditions, of an automatic sleep monitoring system are presented. Data are recorded from patients who were undergoing heart and great vessel surgery. This study resulted in cap, electrode, and preamplifier improvements. Children were used to test the sleep analyzer and medical console write out units. From these data, an automatic voltage control circuit for the analyzer was developed. A special circuitry for obviating the possibility of incorrect sleep staging due to the presence of a movement artifact was also developed as a result of the study.

Read the paper · More papers on PaperTik