Signal processing for hearing impairment.
Harry N. Levitt, Matthew H. Bakke, James Mitchell Kates, Arlene C. Neuman, Teresa Schwander, Mark R. Weiss · PubMed · 1993
Four noise reduction methods for use in sensory aids for hearing impairment were evaluated. These include a two-microphone adaptive noise canceller, short-term Wiener filtering, a transformed spectrum subtraction technique, and sinusoidal modelling. The largest improvements in speech recognition were obtained with the two-microphone adaptive noise canceller in a moderately reverberant room. Significant improvements were also obtained for short-term Wiener filtering for some hearing-impaired subjects. The transformed spectrum-subtraction technique failed to improve performance as the front-end of a hearing aid, but yielded improvements in performance as a preprocessor for the Nucleus Cochlear Implant. Sinusoidal modelling resulted in significant improvements in signal-to-noise ratio, but without a corresponding improvement in speech intelligibility.