Multiresolution Analysis Of Transient Evoked Otoacoustic Emission
Yousef Salimpour, Mohammad Djavad Abolhassani, Alireza Ahmadian, Kamran Barin · 2005
To improve transient evoked otoacoustic emission (TEOAE) test accuracy, a new technique in presented based on multiresolution signal decomposition. Using a wavelet function which optimized according to an active cochlear model, time frequency distribution of TEOAE is obtained. The coefficients of the filter bank are determined from Morlet mother wavelet function in the analysis section. In the synthesis section, time windows are obtained from deduced latency curve. By applying this multiresolution analysis, the correlative noise would be canceled. Therefore, the wave reproducibility can be increased without loss of measuring reliability. In abnormal cases where response is not expected the correlative noise is considered as response are completely disappeared. Thus this will increase the accuracy of screening results.