A Simple Transistorized Linear Scale Ohmmeter
Anwar A. Khan · IEEE Transactions on Instrumentation and Measurement · 1974
The detailed design and the analysis of a simple, accurate, and inexpensive linear scale ohmmeter are presented. The accuracy of measurement by the new ohmmeter is within ±1 percent in the range 10-1-106 Ω, ±2 percent at 10 MΩ, and ±10 percent at 100 MΩ. Unlike most other conventional megohmmeters, this operates with low voltage across the sample and does not require the use of a high input impedance voltmeter. The new circuit exhibits good operational stability.