A Calibration Method for the Logging Instrument Used in Harsh Environments

Zhen Huang, Li Kong, Pu Zhang, Xiao Nan Liu, Cheng Zhang · Applied Mechanics and Materials · 2013

Due to the impact of harsh environments with temperatures higher than 150 °C, the characteristics of the amplifying circuits of logging instrument can be changed, to cause the large measurement error. Therefore, a new calibration method is proposed to obtain the gain value and offset for pre-amplifier system. This method use multiple calibration signals and based on the last square theory to complete the calibration work. At the experiment of calibration for the multi-parameter formation water logging instrument, the gain error for the pre-amplifier system by this calibration be improve to no exceed 0.15% to compare with over 1% by tradition method.