A sliding-band dynamic range compression for use in hearing aids

Nitya Tiwari, Prem C. Pandey · 2014

Sensorineural hearing loss is associated with elevated hearing thresholds, reduced dynamic range, and loudness recruitment. Dynamic range compression in the hearing aids is provided for restoring normal loudness of low level sounds without making the high level sounds uncomfortably loud. A sliding-band compression is presented for significantly reducing the temporal and spectral distortions generally associated with the currently used single and multiband compression techniques. It uses a frequency-dependent gain function calculated on the basis of critical bandwidth based short-time power spectrum and the specified hearing thresholds, compression ratios, and attack and release times. It is realized using FFT-based analysis-synthesis and can be integrated with other FFT-based signal processing in hearing aids to save computation. The technique is implemented and tested for satisfactory real-time operation, with sampling frequency of 10 kHz, window length of 25.6 ms with 75% overlap on a 16-bit fixed-point DSP processor with on-chip FFT hardware.