A 1000 A/20 kV/25 kHz-500 kHz Volt-Ampere-Wattmeter for loads with power factors from 0.001 to 1.00

Kenneth K. Clarke, D.T. Hess · IEEE Transactions on Instrumentation and Measurement · 1996

Many industrial processes and scientific experiments utilize large amounts of ac power at frequencies from 3 kHz to 500 kHz. The phase angle and the impedances of these loads often vary over a wide range. This paper describes an instrument to provide an accurate measurement of currents (1 A to 1000 A), voltages (100 V to 20 kV), and powers (100 W to 20 MW) over the frequency range from 25 kHz to 500 kHz. It deals with loads having power factors down to nearly zero and with load impedances from 10 /spl Omega/ to 20 k/spl Omega/. The paper contains practical details of both the design and the calibration of the front-end voltage and current transducers. The characteristics of a nearly ideal broadband current transducer are presented. Overall instrument calibration, verification and traceability problems are considered in detail.