Determining the influence of Time Constants in Maxwell Bridge
Yan Rong Yang, Qing Lu, Lu Huang, Wen‐Jun Lu · 2014
This paper analyses the self-inductance measurement errors introduced by the time constants of the standard resistors and conductance used in Maxwell bridge, and proposes a common method to measure their time constants. After compensating, in the ranges from 1 mH to 100 H, the self-inductance measurement errors introduced by the time constants are less than 1 × 10-6at 1 kHz.