Measurement of ultrasonic velocity with improved accuracy in pulse echo setup

S. Rajagopalan, Sahil Sharma, P. K. Dubey · Review of Scientific Instruments · 2007

In the present work a personal computer based circuit adopted in pulseecho setup has been developed for the measurement of time-of-flight between the two successive echoes using a differential time measurement with a 16 bit counter and an external clock frequency of 32 MHz. A control program is developed in C to display the time of flight and wave propagation velocity on a user screen. The technique, so developed, minimizes the errors in time delay measurements due to the variation in threshold trigger points of echoes and improves the ultrasonic velocity measurement.

Read the paper · More papers on PaperTik